Plate reader¶
General tips for OD600 measurements¶
- Use flat bottom 96-well plates; round-bottoms might scatter the light
- Use a lid so that your culture does not evaporate. For long measurement sessions, coat the plate's lid with 20% ethanol + .05% Triton X-100 in order to prevent vapor from condensing on the lid and distorting measurements.
- Place 100 uL of a blank medium and 100 uL of each of your samples.
- Samples should start from OD .05-.1.
Using CLARIOstar Plus¶
- Turn on CLARIOstar Plus and open the associated program.
- Create a new protocol or copy an existing one.
- Right click on a protocol to edit.
- If incubation is needed, set it in the program (not protocol).
- Insert the plate in the purple tray (click the small purple button on the device or in the program to open and close the tray).
- Click "Start measurement" in your protocol.
- The device start working and most buttons in the program become disabled.
- Click on "Current State" to see the data as it is accumulating.
- You may pause your run, take out the plate, add something to it, put it back and continue measurements.
- You can also stop the run prematurely. It will ask if you want your data saved. Choose "Yes".
- Open data analysis program (MARS) and export data to csv.
OD600 protocol¶
- In the program, set Temperature: 37C, on.
- Wavelength settings:
- Discrete wavelengths
- No. of wavelengths: 1
- Wavelength: 600 nm
- Pathlength correction: Off
- Well scan:
- Spiral average
- Diameter: 4 mm
- General settings:
- Setting time: 0.1 s
- No.of kinetic windows: 1
- Kinetic window 1:
- No. of cycles: 300
- No. of flashes: 8 (how many measurements to make at each time point)
- Cycle time: 300 s (how often you make measurements)
Using Tecan Spark¶
- Turn on Tecan Spark and the external cooling device.
- If the cooling device has a "Low fluid" indicator on, notify facility manager.
- Open Tecan SparkControl application and choose your protocol or create new.
- In needed, set temperature in the menu bar.
- It takes about 40 min to go from ambient.
- You may also include temperature regulation in the protocol.
- Typical protocol layout:
- Choose plate layout and wells that are measured.
- A custom plate layout can be created under Settings.
- Add kinetic loop that dictates which conditions to repeat and how often.
- Add measurements: Absorbance, Luminescence, Fluorescence.
- Add shaking
- For instance, choose continuous shaking to keep shaking between measurements.
- Choose plate layout and wells that are measured.
- Save the protocol.
- Hit start.
- You can observe measurements in real time but often not for all wells.
- If you hit Stop, the run will stop immediately.
- Data is exported automatically into an Excel spreadsheet once the run is over.
- You can change the settings how data is exported by going to Settings > Data Handling.
- There is no option to reexport data in a different format after the run is initiated (or done).
- Turn of Tecan Spark and the external cooling device.
OD600 protocol¶
Kinetic (loop):
- Kinetic cycles: 300
- Interval time: 00:05:00
Within each cycle: - Absorbance: - Measurement wavelength: 600 nm - Number of flashes: 1 - Settle time: 100 ms - Multiple Reads per Well: 3x3 Square (filled), Border 1000 μm - Shaking: - Pattern: Double Orbital - Continuous - Position (means whether to measure at the incubation chamber or just naywhere where the plate is at the time): Incubation - Amplitude: 1 mm - Frequency: 270 rpm (automatically set based on amplitude)