Styx¶
Styx is used to sort droplets based on fluorescence.
Procedure¶
Good practice to check under the microscope first to see if there is any fluorescence.
- Arrangement:
- Spacing oil – usually a large quantity (e.g., in a 5 mL Eppendorf or a 25 mL tube), tubing need not do all the way down to the bottom.
- Sample – insert the tubing all the way down and as you keep screwing gradually pull it out such that it's at the very bottom but not touching the tube.
- Empty
- Waste – no pressure control or screws needed
- Positive selection – no pressure control or screws needed
- Purging – use Start button:
- Spacing oil: 3 kPa – wait until it fills the chip
- Sample: 1 kPa
- Do not plug into the chip yet.
- Wait for the spacing oil to fill in while holding the tubing upwards (so that the liquid doesn't go out).
- Once the oil is in, lower the tubing into a clean Eppendorf and wait until excess oil is fully drained an a couple of drops with droplets accumulate at the tip of the tubing.
- Plug into the chip.
- Running:
- Spacing oil: 12 kPa (or 21 kPa)
- Sample: 8 kPa (or 15 kPa)
- Voltage: 700-800 V – for best quality, make sure the sorted droplet goes into its outlet without touching anything; can also adjust pulse duration to 3000 us and pulse delay to 0 us
- Laser: align to the right tip of the electrode
- Start the laser(s) of the desired wavelength
- Start fluorescence data collect (in the fluorescence tab)
- Start voltage
- Clean up:
- Take out the tubing of positive and waste and hold vertically for the droplets to drain into the tubes
- Stop the experiment (voltage, laser, and fluorescence detector)
- Disconnect everything